Installing the Heatsink/Measuring Cell Assembly

!
Ensure there is no power being supplied to the analyzer during the
installation of the Heatsink/Measuring Cell Assembly.
WARNING
!
While installing the Heatsink/Measuring Cell Assembly, take care not
to damage the joining surfaces (flamepath) of the Housing Seal on the
CAUTION
Ex Housing and Heatsink/Measuring Cell Assemblies.
To install the Heatsink/Measuring Cell Assembly:
1. Connect the Heatsink/Measuring Cell Assembly to the Housing Seal
on the Ex Housing (Figure 3-2), being careful not to damage the pins
on the connector.
2. Connect the tube from the Pressure Regulator to the fitting on the
Heatsink/Measuring Cell Assembly that is closest to the backpan.
Connect the tube from the Filterblock Assembly to the fitting on the
Heatsink/Measuring Cell Assembly that is farthest from the backpan.
Installation and Start-Up
|
3-9