6-36 | Model 241CE II Hydrocarbon Dewpoint Analyzer
Clearing (Resetting) Alarms
Active alarms can be cleared (reset) from the Stat\History menu. Clear-
ing the alarms removes any active alarms from the User Interface display
and stores them (up to a maximum of nine) in the History Buffer (^SC
or ^XSC), also in the Stat\History menu. A password is not required to
perform this command.
To clear alarms:
1. From the RUN menu, press to view the Stat\History menu. The
first menu item you will see is the "En\Dis Alrm?" prompt.
2. Press once to view the "Clear Alarm?" prompt. While this prompt
is on the User Interface, press the '+' button to clear the alarms; the
message "Cleared" will appear on the User Interface.
If you do not wish to clear the alarms, press the '–' button; the mes-
sage "NotCleared" will appear on the screen. Or, you can press to
scroll past the "Clear Alarm?" prompt and leave the alarms in their
current state.
Enabling/Disabling the Off-Spec Alarm
The Off-Spec Alarm can be enabled (turned on) or disabled (turned off)
from the Stat\History menu. When the Off-Spec Alarm is enabled, LED 1
is Off; when the Off-Spec Alarm is disabled, LED 1 is On to alert the op-
erator to this condition. A password is not required to perform this com-
mand.
To enable or disable the Off-Spec Alarm:
1. From the RUN menu, press to view the Stat\History menu. The
first sub-menu item you will see is the "En\Dis Alrm?" prompt.
2. If the Off-Spec Alarm is already in the current state you require, press
to scroll past the "En\Dis Alrm?" prompt and leave this parameter
in its current state.
To enable the Off-Spec Alarm if it is currently disabled, press the '+'
button. LED 1 will turn Off to indicate the Off-Spec Alarm is enabled.
To disable the Off-Spec Alarm if it is currently enabled, press the '–'
button. LED 1 will turn On to indicate the Off-Spec Alarm is disabled.