6-42 | Model 241CE II Hydrocarbon Dewpoint Analyzer

Status Code (SCode), Alarm Conditions, and Corrective Action

This section lists the primary code, error code, alarm type, condition/
description, and suggested corrective action associated with each type of
SCode alarm. If no alarms are present, the primary SCode '0' is displayed.
The 'w' or 'f ' preceding a message indicates a 'w'arning or 'f 'ault
alarm condition.
NOTE
Alarm
Type
1
Warning
2
Warning
Condition / Description / Suggested Corrective Action
No Warnings or Faults
If the primary code returned is '0', no Warnings or Faults exist.
This is not an alarm.
Corrective Action:
No action is required.
w TimeOut [Stage-Time-Out]
One or more of the stages in the measuring cycle has
encountered a time-out condition.
Corrective Action:
Review the current XSCode (Stat\History menu) to
determine which stage(s) have a time-out condition
– "w Hold TimeOut", "w Cool TimeOut", and/or
"w Warm TimeOut". Depending on which stage(s) have
encountered this condition, refer to its specific corrective
action.
w No H2O DP [Water-Dewpoint-Temperature-Not-Found]
This alarm is available only if the H2O DP:INT option is
enabled from the MonitorCFG menu.
The internal water dewpoint temperature was not determined
during the last measuring cycle. This alarm is not related
in any way to the optional, external water dewpoint
temperature (if used).
Corrective Action:
Review the current XSCode (Stat\History menu) to
determine if there has been a "w Cool TimeOut" alarm.
If not, decrease the mirror low setpoint temperature
(ApplicCFG : MTLo) and check the HCDP temperature
(RUN menu) to see if it is particularly high (e.g., greater
than 20 °C/68 °F).