6-46 | Model 241CE II Hydrocarbon Dewpoint Analyzer
Extended Status Code (XSCode), Alarm Conditions, and
Corrective Action
This section
lists the primary code, error code, alarm type, condition/de-
scription, and suggested corrective actions. If no alarms are present, the
primary XSCode '0' is returned by the system.
The 'w' or 'f ' preceding a message indicates a 'w'arning or 'f 'ault
alarm condition.
NOTE
Alarm
Type
1
Warning
Condition / Description / Suggested Corrective Action
No Warnings or Faults
If the primary code returned is '0', no Warnings or Faults exist.
This is not an alarm.
Corrective Action:
No action is required.
w EEP Warn [End-Of-EEPROM-Life]
The number of times that configuration data has been rewritten
to the EEPROM has exceeded the safe number.
Corrective Action:
Replace the EEPROM as soon as possible. Record all
configuration values before removing the EEPROM. Upon
receiving a new EEPROM from AMETEK, re-enter the
parameters to ensure your analyzer will operate the same
as it did prior to replacing the EEPROM.
Alternatively, you can send the configuration values and
P.O. number for the EEPROM to AMETEK at the same
time. At the factory, AMETEK will program the customer-
specific parameters in the EEPROM. In this case, you will
only have to install the EEPROM; the parameters will take
effect immediately upon power-up.
After installing the new EEPROM, enter the configuration
values you recorded before removing it (if the EEPROM
was not pre-configured by AMETEK).