Sub-Menu Item
Tz
^WDz
^HCDz
^WCz
^WMirz n
^HCMirz n
^SCz n
^XSCz n
EEP WR n
000000000-1
M241 Vx-xx
4-14 | Model 241CE II Hydrocarbon Dewpoint Analyzer
Description
This system status parameter (T) indicates the Run-Time (Days Hours:Minutes – 000 00:00) at the
beginning of the Cooling Stage for each of the last nine measuring cycles (where 'z' = 1–9: '1' = most
recent entry and '9' = oldest entry).
This history buffer displays the Water Dewpoint temperature (°C or °F) for the last nine measuring
cycles (where 'z' = 1–9: '1' = most recent entry and '9' = oldest entry).
Note: This signal will be displayed only if an external water signal is provided and if
the H2O DP:EXT option is enabled from the MonitorCFG menu. If the H2O
DP:INT option is enabled, the WDP readings are derived from the internal
water mirror surface – not recommended.
This history buffer displays the Hydrocarbon Dewpoint temperature (°C or °F) for the last nine mea-
suring cycles (where 'z' = 1–9: '1' = most recent entry and '9' = oldest entry).
This history buffer displays the external Water Content (optional, if used) in mg/m
or PPMW values for the last nine measuring cycles (where 'z' = 1–9: '1' = most recent entry and '9' =
oldest entry).
Note: This signal will be displayed only if an external water signal is provided and if
the H2O CT:EXT option is enabled from the MonitorCFG menu. If the H2O
DP:INT (with water content) option is enabled, these readings are derived from
the internal water mirror surface – not recommended.
This history buffer displays the phototransistor signal from the Water-side of the mirror at the start of
the Run cycle, for the last nine measuring cycles (where 'z' = 1–9: '1' = most recent entry and '9' =
oldest entry, and where 'n' = the displayed phototransistor signal value).
This history buffer displays the phototransistor signal from the Hydrocarbon-side of the mirror at the
start of the Run cycle, for the last nine measuring cycles (where 'z' = 1–9: '1' = most recent entry and
'9' = oldest entry, and where 'n' = the displayed phototransistor signal value).
This history buffer displays the basic Status Code alarms for the last nine measuring cycles (where
'z' = 1–9: '1' = most recent entry and '9' = oldest entry, and where 'n' = the total sum of all primary
SCode alarms triggered during the last measuring cycle).
This history buffer displays the basic Extended Status Code alarms for the last nine measuring cycles
(where 'z' = 1–9: '1' = most recent entry and '9' = oldest entry, and where 'n' = the total sum of all
primary XSCode alarms triggered during the last measuring cycle).
This system status parameter displays the number of writes ('n' = total number of writes) to the ana-
lyzer's EEPROM (changes to the system configuration settings).
This system status parameter displays the Serial Number of the analyzer.
This system status parameter displays the current Firmware Version running the Model 241CE II
Analyzer.
3
, lb/MMcf, PPMV,