Sub-Menu Item
Description
I3Z
V/I Analog output #3 Zero calibration value, WCT (4.00 mA).
Note:
This parameter will be displayed only if an external water signal is provided
and if the H2O CT:EXT option is enabled from the MonitorCFG menu.
I3S
V/I Analog output #3 Span calibration value, WCT (20.00 mA).
Note:
This parameter will be displayed only if an external water signal is provided
and if the H2O CT:EXT option is enabled from the MonitorCFG menu.
I4Z
V/I Analog output #4 Zero calibration value, Status Code (4.00 mA).
I4S
V/I Analog output #4 Span calibration value, Status Code (20.00 mA).
Note: See "Current Output Zero/Span Calibration," following this section, for informa-
tion on how to calibrate the Zero/Span V/I outputs.
This prompt asks if you wish to change Passcode2. To change the password for the MonitorCFG
Chng PSCD2?
menu, enter a new 4-character code, using any combination of the , '+', and '–' buttons. After
entering the password, "Chng PSCD2?" will be displayed again. Enter the password a second time.
Save MonCFG?
This prompt asks if you wish to permanently save the changes you made to the MonitorCFG
parameters. To save any changes you made, press the '+' button. If the changes are not saved, they
will remain in effect until the analyzer is powered down or reset.
|
User Interface
4-21