SYSTEM OVERVIEW
Positioning Laser
Press the laser trigger to activate the positioning laser.
The positioning laser is marked with the words "LASER APERTURE".
Warning
Do not look directly into the laser-emitting aperture. The Class 2 laser module of the
camera emits visible radiation that may be harmful to the human eye. Looking directly at
the laser diode emission at close range can cause eye damage.
Warning
Take precautions to ensure that people with restricted movement or reflexes (for example,
patients undergoing medical procedures) do not look directly into the laser-emitting
aperture. Patients undergoing medical procedures may lack protective reflexes (turning
away eyes and/or head, closing eyes) due to pharmaceutical influences and/or mechanical
restraints. The Class 2 laser module on the Position Sensor emits radiation that is visible
and may be harmful to the human eye. Direct viewing of the laser diode emission at close
range may cause eye damage.
The camera is a highly sensitive optical device. Handle it with care.
CM Camera LEDs and Acoustic Signals
No.
LED Indications
The CM Camera requires a thermal stabilization period in order to provide reliable measurements.
When the Position Sensor is powered on, the power light will flash to indicate that the system is
warming up. When the light stops flashing, the system is ready for use.
System and Technical User Guide Rev. 1.2 Buzz Navigation (Ceiling-Mounted) Ver. 1.0
Component
Power LED
Error LED
Figure 6
27