COMPLIANCES AND SPECIFICATIONS
4.2
Environmental Conditions
Local Restrictions
• Store and operate the system in locations that are protected against moisture, wind, sunlight,
dust, salinity and sulfur.
• Do not store system in the close vicinity of chemical products or gas.
• Do not expose system to direct UV light.
• Exposing the device to environmental stress above or below the specified boundaries (e.g.,
fumigation for pest control) may lead to damage or reduced lifetime of the device.
Altitude Considerations
• Unless stated otherwise, the system is rated for use at an altitude less than 3000 meters.
• Transport or store the system at an altitude less than 6000 meters.
Adaption Time
Adaption time for use after extreme storage conditions is a minimum of 24 hours.
Temperature, Humidity and Pressure Conditions
Specification
Ambient Temperature
Relative Humidity
Pressure
System and Technical User Guide Rev. 1.2 Buzz Navigation (Ceiling-Mounted) Ver. 1.0
Operating Conditions
For components inside the op-
erating room:
10°C to 30°C [50°F to 86°F]
For components outside the
operating room:
10°C to 35°C [50°F to 95°F]
30% to 75% (non-condensing)
700 to 1060 hPa
Transport/Storage Conditions
-10°C to 45°C [14°F to 113°F]
10% to 90% (non-condensing)
500 to 1060 hPa