Mainstream Co - Dräger Infinity M540 Instructions For Use Manual

Acute care system
Hide thumbs Also See for Infinity M540:
Table of Contents

Advertisement

Mainstream CO

Priority
Message
None
CO
calibration check
2
failed
None
CO
calibration check in
2
progress
None
CO
calibration check
2
successful
None
CO
calibration failed
2
None
CO
calibration in
2
progress
None
CO
calibration
2
successful
!
%0 check airway
1)
adapter
!
%0 H/W failure
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
2
Cause
The mainstream sensor
calibration procedure failed.
The mainstream sensor
calibration procedure is in
progress.
The mainstream sensor
calibration procedure was
successful.
The mainstream sensor
calibration procedure was
unsuccessful.
The mainstream sensor
calibration procedure is in
progress.
The mainstream sensor
calibration procedure was
successful.
– The mainstream sensor
is not properly seated on
the adapter
– There are secretions in
the adapter.
– There is sensor zero drift
1)
CO
2
failure.
sensor hardware
Troubleshooting
Remedy
Contact DrägerService
Informational message – no
action required.
Informational message – no
action required.
Try again or contact
®
DrägerService
.
Informational message – no
action required.
Informational message – no
action required.
– Make sure the mainstream
sensor is attached properly
to the adapter.
– If message persists, clean
or replace the airway
adapter.
– If message persists though
the airway adapter is clean,
zero the sensor.
Contact DrägerService
®
.
®
.
543