Configuring the auto view settings
You can perform various functions in the Auto view
page. The following settings describe the general
settings of the Auto view page. For detailed
information on setting up the display attributes of a
parameter, see "Configuring parameters for
display" on page 448.
To configure the available settings
In the following steps, the letters in parentheses
correspond to the diagram for the Auto view page
(see page 446).
1 Access the Auto view page (see page 444).
2 Select the display mode by selecting one of the
following two buttons next to Display mode:
– Auto (A) to select the auto display mode
(see page 74).
– Manual (B) to select the manual display
mode (see page 68).
3 Select the Waveforms button (E) to determine
the number of waveforms that can be selected
in the parameter selection window (N). The
number of waveforms available for selection
depends on the purchased software and
hardware options:
Device With option
C500
12 or 16 waveforms 10 waveforms
C700
16 waveforms
Select the Layout button (F). Then select the
Left or Right (default) button to determine if the
waveforms appear to the left or to the right of
the parameter fields.
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Without option
12 waveforms
4 Select the Pressure overlap on or off (default)
button (G) to activate or deactivate pressure
overlap mode. This feature works only if the
pressure waveforms are displayed in adjacent
channels.
5 Select the Parameter boxes button (H) and
use the rotary knob to select the desired
number of parameter fields for display. The
available selections are: Off, 1, 2, 3 (default), 4,
5, 6.
6 Select the Layout button (I). Then select the
Top or Bottom (default) button to determine if
the parameter fields appear along the bottom or
the top of the screen.
– Select the Split screen button (J). This button
appears grayed out if the web enabled layouts
option is locked. The available selections are:
None (default), Anesthesia show all, BIS
show all, BIS/NMT show all, CCO show all,
ECG/ST, ECG/Vent, ECG show all, Loops,
SpO2 show all, ST parameters, Multi-tab
split screen,Ventilator show all, Trend table.
If you select None, the monitoring area contains
only real-time parameters. Any other selection
divides the monitoring area into two windows.
The right window continues to display the real-
time parameters. For more detail see "M540
and Cockpit communication" on page 46.
7 Select the Mini trends button (K) to activate or
deactivate the mini-trend display or select a
trend display time (see page 72). The available
selections are: Off, 10 min, 15 min, 20 min,
30 min (default), 45 min, 1 h, 90 min, 2 h, and
4 h.
8 Select the NIBP trend button (L) to choose
between the graphic or numeric representation
of the NIBP mini-trend display.
9 Toggle the Toolbar button (M) to On (default) or
Off to activate or deactivate the auto view setup
toolbar (see page 447).