System configuration
Biomed setup – service setup
The following table lists the available settings of the
Service page. To access this page, see page 470.
System setup
B
H
478
A
C
D
E
F
G
I
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
A Biomed tab
B Product identification field displaying software-
specific information (for example, software
version, and so on).
C Select this button to copy all logs to the
connected USB flash drive.
D Select this button to export the current profiles
to the connected USB flash drive.
E Select this button to import the profiles from the
connected USB flash drive.
F Select this button to export the shared system
profile
G Select this button to import the shared system
profile
H Window displaying status messages relating to
the function that is being executed.
I
Service button for accessing Service-related
functions such as unlocking options (refer to the
Technical documentation available from
DrägerService for detailed information).