Alarms
Changing alarm settings for a single parameter
In the steps below, the letters in parentheses refer
to the diagram of the parameter-specific setup
page.
WARNING
Setting alarm limits to extreme values may
prevent certain alarm conditions from being
detected and from being annunciated with
acoustic and optical alarm signals.
To configure the alarm settings
1 Select Sensor parameters... on the main menu
bar.
2 Select the desired parameter tab (for example,
ECG).
or
 Select the parameter field to access the
parameter setup page directly.
128
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
3 Select the Alarm on/off button (A), to activate or
deactivate alarm monitoring. A crossed-out
triangle appears in the parameter field when
alarm monitoring is deactivated.
NOTE
If French NFC mode is activated (see page 473),
you cannot deactivate HR alarms.
4 Select the setup button (C) to adjust the upper
alarm limits.
5 Select the setup button (D) to adjust the lower
alarm limits.
6 Select one of the following settings for the
Archive buttons (E) to determine what happens
in response to an alarm:
– Off – no event is stored and no recording is
generated.
– Store – stores the event for later review
(see page 140).
– Record – generates a timed recording.
– Str/Rec – stores an event for later review
and generates a timed recording.