External Device Disconnection Alarm; The Code Function - Dräger Infinity M540 Instructions For Use Manual

Acute care system
Hide thumbs Also See for Infinity M540:
Table of Contents

Advertisement

External device disconnection alarm

If the external device alarm feature is activated at
the Cockpit (see page 461) and an external device
is disconnected from the Cockpit, the following
happens at the ICS and at the Cockpit:

The Code function

You can configure a set of monitoring functions that
can be activated during emergency care by
selecting the Code button on the main menu bar.
Depending on which of these settings are activated
(see page 464), any of the following happens when
you select the Code button:
– A continuous recording starts
– NIBP measurements start in continuous mode
– The alarm volume of the alarm condition with
the highest priority is automatically reduced to
the minimum setting.
NOTE
When the Audio off setting is set to Off (see
page 461), the message Audio off and the
symbol
appear in the Cockpit header bar
when you invoke the Code function.
– The pre-configured All alarms off setting
determines if the alarm annunciation is
deactivated and the All alarms off message
appears in the header bar when you press the
Code button (see page 465).
NOTE
The All alarms off button works independently of
the pre-configured Code setting. You can
therefore activate or deactivate the All alarms off
button even after pressing the Code key.
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
– An alarm tone of low priority sounds.
– The message External device disconnected
appears.
In addition to activating the pre-configured features,
a timer with a red background appears in the
header bar with the following two buttons:
– Stop for stopping the timer. The label of the
button changes to Start.
– Reset button for resetting the timer to zero.
The Code button does not function unless the
M540 is docked.
NOTE
When the All alarms off setting is set to Off (see
page 461), the message Audio alarms off and
the
symbol appear in the Cockpit header bar
when you invoke the Code function.
To activate the Code function
 Press the Code button on the main menu bar.
To deactivate the Code function
 Press the Code button on the main menu bar a
second time. All functions are deactivated.