Dräger Infinity M540 Instructions For Use Manual page 153

Acute care system
Hide thumbs Also See for Infinity M540:
Table of Contents

Advertisement

Parameter
Alarm
default
status
IBP S pediatric/
neonate
Increment:
(GP1 S to
1 mmHg or
GP4 S,
0.1 kPa
LV S,
RV S)
On
(ART S,
PA S)
IBP D adult
Increment:
1 mmHg
(GP1 D to
or 0.1 kPa
GP4 D,
LV D,
RV D)
On
(ART D,
PA D)
IBP D pediatric
Increment:
1 mmHg
(GP1 D to
or 0.1 kPa
GP4 D,
LV D,
RV D)
On
(ART D,
PA D)
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Alarm limit range
Upper: –24 to +300 mmHg
–3.2 to +40.0 kPa
Lower: –25 to +299 mmHg
–3.3 to +39.9 kPa
Upper: –24 to +300 mmHg
–3.2 to +40.0 kPa
Lower: –25 to +299 mmHg
–3.3 to +39.9 kPa
Upper: –24 to +300 mmHg
–3.2 to +40.0 kPa
Lower: –25 to +299 mmHg
–3.3 to +39.9 kPa
Upper limit
Lower limit
defaults
defaults
– 120 mmHg
– 50 mmHg
(16.0 kPa)
for GP1 S to
GP4 S,
ART S, LV S
– 35 mmHg
(4.7 kPa) for
– 75 mmHg
PA S, RV S
– 10 mmHg
– 110 mmHg
– 50 mmHg
(14.7 kPa)
for GP1 D to
GP4 D,
ART D
– 25 mmHg
(3.3 kPa) for
– 2 mmHg
LV D
– 13 mmHg
(1.7 kPa) for
PA D, RV D
– 80 mmHg
– 30 mmHg
(10.7 kPa)
for GP1 D to
GP4 D,
ART D
– 25 mmHg
(3.3 kPa) for
– 2 mmHg
LV D
– 13 mmHg
(1.7 kPa) for
PA D, RV D
Alarms
Archive
default
setting
Off
(6.7 kPa)
for GP1 S
to
GP4 S,
ART S
(10.0 kPa)
for LV S
(1.3 kPa)
for PA S,
RV S
Off
(6.7 kPa)
for GP1 D
to
GP4 D,
ART D
(0.3 kPa)
for PA D,
LV D,
RV D
Off
(4.0 kPa)
for GP1 D
to
GP4 D,
ART D
(0.3 kPa)
for PA D,
LV D,
RV D
153