System configuration
Configuring the patient-specific profiles
In the following steps, the letters in parentheses
correspond to the diagram for the patient-specific
Profiles page (see page 487).
To configure the profiles
1 Access the patient-specific profiles page.
2 Select the desired profile in the pick list (K).
3 Select one or more of the following buttons:
– Select the Set as Default button (E) to
designate the selected profile as the new
default profile for the selected patient
category. After each patient discharge or a
restart of the Cockpit, the default profile is
automatically loaded when that patient
category is selected.
– Select the Delete Profile button (F) to
delete the selected profile.
488
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
– Use the
button next to the Profile name
field (G) of the profile page to name the
profile. The name appears in the Select
profile page (see page 484).
– Use the
button next to the Description
field (H) of the profile page to add or modify
an existing description. The description
appears in the Select profile page (see
page 484) when you select a profile.
– Select the arrow button next to the Default
view field (I) and select the view that you
wish to designate as the default view.