Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Available settings
– 1, 2 (default), 3, 4, 5 min
– No timeout
– Disabled
On (default), Off
Description
The button on the alarm toolbar changes to All
alarms paused. This button is accessible by
selecting the
symbol on the quick access
toolbar, (see page 48).
When the button is selected, the following
happens:
– All alarm functions are temporarily
suppressed for the selected time. The alarm
function is automatically activated when the
alarm pause timer times out.
– The message All alarms paused appears
in the header bar on yellow background with
a timer and the following symbol:
The button on the alarm toolbar changes to All
alarms off. This button is accessible by
selecting the
symbol on the quick access
toolbar, (see page 74).
When the All alarms off button is selected, the
following happens:
– All alarm functions are suppressed until you
select the button again which activates the
alarm function.
– The message All alarms off appears in the
header bar on red background with the
following symbol:
The All alarms paused button on the alarm
toolbar is grayed out and you cannot
temporarily or permanently deactivate alarm
monitoring.
When this function is activated, alarm
conditions are verified for a certain time before
triggering acoustic and optical alarm signals
(see page 111). This feature reduces nuisance
alarms.