Connecting the Masimo rainbow SET MCable
The Masimo rainbow SET MCable connects
directly to the M540. The logo on the MCable
identifies if you are using a Masimo rainbow SET or
a Masimo SET MCable.
A color band located on the side of the Masimo
rainbow SET MCable indicates which parameters
are activated.
– Fields appearing in color represent parameters
that are already activated
– Fields with the letter 'X' denote parameters that
are not activated
– Fields that appear empty denote parameters
that might be activated later
A Masimo MCable can be mounted to the back of
an M540 (see page 97).
To connect the Masimo rainbow SET MCable
1 Attach the MCable connector (B) to the blue
SpO2 port (A) of the M540.
2 Attach the intermediate cable (D, E) to the 20-
pin connector of the MCable (C).
3 Attach the appropriate Masimo sensor to the
end of the intermediate cable (F). For detailed
information on which sensors support which
parameters, refer to the instructions for use
Infinity Acute Care System – Monitoring
Accessories.
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
SpO2 and Pulse CO-Ox monitoring with Masimo SET MCable
A SpO2 port on the M540
B MCable connector
C MCable (20-pin connector)
D Masimo rainbow SET intermediate cable
(connector to MCable)
E LNCS intermediate cable (connector to

MCable)

F Connector for various sensors