Parameter
Alarm
default
status
inCO2
On
Increment of
1 mmHg,
0.1 kPa, or 0.1
etCO2
On
Increment of
1 mmHg,
0.1 kPa, or 0.1%
PVC/min
On
Increment of 1
inHal
On
Increment of
0.1 kPa or 0.1%
etHal
Increment of
0.1 kPa or 0.1%
inIso
On
Increment of
0.1 kPa or 0.1%
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Alarm limit range
Upper: 2 to 10 mmHg
0.3 to 1.3 kPa
0.3 to 1.3%
Lower: not user-selectable
Upper: 6 to 99 mmHg
0.8 to 13.3 kPa
0.8 to 13.3%
Lower: 5 to 99 mmHg
0.7 to 13.2 kPa
0.7 to 13.2%
Upper: 1 to 50
Upper: 0.1 to 8.5 kPa
0.1 to 8.5%
Lower: 0.0 to 8.4 kPa
0.0 to 8.4%inHal
Upper: 0.1 to 8.5 kPa
0.1 to 8.5%
Lower: 0.0 to 8.4 kPa
0.0 to 8.4%
Upper: 0.1 to 8.5 kPa
0.1 to 8.5%
Lower: 0.0 to 8.4 kPa
0.0 to 8.4%
Upper limit
Lower limit
defaults
defaults
4 mmHg
Not applicable Off
(0.5 kPa, 0.5%)
50 mmHg
30 mmHg
(6.7 kPa, 6.6%)
(4.0 kPa,
3.9%)
10
Not applicable Off
1.6 kPa, 1.6%
0.0 kPa, 0.0% Store
(adult)
1.9 kPa, 1.9%
(pediatric)
8.5 kPa, 8.5%
0.0 kPa, 0.0% Off
2.4 kPa, 2.4%
0.0 kPa, 0.0% Store
(adult)
2.8 kPa, 2.8%
(pediatric)
Alarms
Archive
default
setting
Off
157