Co Parameter Setup - Dräger Infinity M540 Instructions For Use Manual

Acute care system
Hide thumbs Also See for Infinity M540:
Table of Contents

Advertisement

Microstream CO
monitoring
2
CO
parameter setup
2
Setup functions for CO
the CO
dialog within the Microstream tab.
2
NOTE
When a Scio module is connected, parameter
controls for CO
are available only in the Scio
2
setup menu.
Selection
RRc apnea time
[s]
RRc apnea
archive
Next service in:
Averaging
Scale
Color
Last calibration:
354
parameters take place in
2
Available settings
Off (default)
10 s, 15 s, 20 s, 25 s, 30 s
Off, Store (default), Str/Rec
Informational only (settings
are not applicable)
Last valid breath,
Off,
10 s, 20 s (default), 30 s
– 0 to 40 mmHg (default), 0
to 60 mmHg,
0 to 80 mmHg, 0 to
100 mmHg
– 0 to 5 kPa (default), 0 to
8 kPa, 0 to 12 kPa, 0 to
16 kPa
– 0 to 5% (default),
0 to 8%, 0 to 12 %,
0 to 16%
Red, White, Yellow (default),
Green, Light blue, Blue,
Purple, Orange
Informational only
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
For detailed alarm setup information, see "Config-
uring the alarm settings for a patient" on page 125.
Description
Specifies how long the Cockpit waits before
reporting a cessation in breathing as an apnea
event.
Determines what happens in response to an
apnea event.
The remaining hours until maintenance is
required.
Controls the specific time or the interval used
to select the maximum measured etCO
the minimum measured inCO
Adjusts the size of the CO
Scale settings apply to both Mainstream and
Microstream.
Determines the color of the CO
and the parameter labels and values.
Displays the date of the last calibration.
and
2
.
2
waveform.
2
waveform,
2