System configuration
Selection
Audio pause: Quiet
mode
Alarm setup – Code functions
For urgent care, you can configure a set of
individual monitoring functions. These functions
can be activated simultaneously when you select
the Code button on the main menu bar. For more
information regarding this function, see page 147.
When the Code button is pressed, a timer along
with a Stop and a Reset button appears in the
header bar.
Selection
Continuous
recording
Continuous NIBP
mode
Audio off
464
Available settings
– On
– Off (default)
Available settings
– On
– Off (default)
– On
– Off (default)
– On
– Off (default)
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Description
– On – Only alarm conditions of high priority
override an active audio pause. The
appropriate parameter field flashes. Alarm
conditions of equal or lower alarm priority
will not be reported with an alarm tone.
– Off – Any new alarm condition, regardless
of its alarm priority, overrides an already
active audio pause state at the Cockpit and
at the ICS if the patient is admitted there. All
optical and acoustic alarm signals are
reported fully for any new alarm condition.
For detailed information how quiet mode affects
the audio pause behavior, see page 121.
The following table lists the available settings of the
Code page. To access this page, see page 458.
Description
– On – a continuous recording starts when
you select the Code button.
– Off – no recording starts when you select
the Code button.
– On – continuous NIBP measurements start
when you select the Code button.
– Off – no NIBP measurements start when
you select the Code button.
– On – No acoustic alarm signal sounds when
you select the Code button.
– Off – Alarm tones sound when when you
select the Code button.