Alarms
Message
1)
SpO2 sensor off
SpO2 check sensor
ECG leads off
1)
Message originating from a Masimo rainbow SET or a Masimo SET MCable.
132
Available settings
Archive settings
– Off (default)
– Store
– Record
– Str/Rec
Alarm settings
Alarm column
– High
– Medium
– Low (default)
– One-shot
(off)
Archive settings
– Off (default)
– Store
– Record
– Str/Rec
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Description
The archive settings cannot be configured for these
two alarm messages. The archive setting follows
the general archive status for the parameter. For
details on how to change the archive settings of a
parameter, refer to page 127.
Determines what happens when the corresponding
alarm occurs: – generates a timed recording and
stores the event.
– High: The event is treated as a latching alarm.
– Medium: The event is treated as a medium-
priority alarm.
– Low: The event is treated as a persistent low-
priority alarm.
– One-shot: The event is treated as a single
notification alarm of low priority. The message
ECG leads off appears briefly in the header bar
until the user acknowledges the condition or the
condition disappears.
: No optical or acoustic alarm signals are
triggered.
These archive settings cannot be configured for this
alarm message. The archive setting follows the
general archive setting for the parameter. For
details on how to change the archive settings of a
parameter, refer to page 127.