Brainlab ExacTrac Planning Manual page 22

With a varian truebeam linear accelerator
Table of Contents

Advertisement

Cabling
#
From
To
Generator cab-
Generator cabinet 2
S
inet 1 (2.0a)
(2.0b)
In-room distri-
Flat panel detectors
bution box
T
(2.3a/2.3b)
(1.3)
Computer cabi-
Hospital network
U
net (1.0)
(4.0)
1)
The cabling from the computer cabinet to the in-room power box and the flat panels can be
placed together in a single 100 mm conduit.
2)
Depending on system configuration, one or two cables are required.
3)
A 130 mm conduit is required from generator 1 to floor box 1 (labeled "a") and from generator 2
to floor box 2 (labeled "b"). If generators 1 and 2 are positioned directly next to each other, it is
possible to place both X-ray tube cables into a single conduit. In this case, separate the high
voltage cables from the data cables.
22
Cable
Cable Di-
Largest Plug Diame-
Length
ameter
ter
5 m (op-
2x 6, 24
tional 20
35 mm
mm
m)
n/a
9 mm
No plug
4 m
5 mm
14 mm
Minimum Conduit Di-
ameter
75 mm (low and high
voltage cables togeth-
er in the same conduit,
no sharp turns)
25 mm (see page
77)
25 mm
Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x