Cabling
Further Table Information
Abbreviations:
• W.: Warning
• c: Mains cables not provided by Brainlab
•
1)
•
2)
•
3)
• DP: Electric Distribution Panel
• PEG: Protective Earth Grounding spot
Additional Cable Information:
• Cable 3: Label cable 3 with "Monitor" on the computer cabinet side.
• Cable 4: The IR USB adapter cable is equipped with a device to combine power and data,
which needs to be stored carefully and kept protected from dust and shocks.
• Cable 8a: Label cable 8a with "Linac" on the computer cabinet side.
• Cables 15, 16 and 17: If you ordered a ceiling-mounted touch screen monitor, cables 15, 16
and 17 come pre-installed in the monitor arm.
• Cable 22: Minimum bend radius is 20 mm.
• Cable 23: Minimum bend radius is 55 mm.
NOTE: In some cases, cable direction is important. Refer to the column, "Note Cable Direction".
26
A total of 60 m of cable is included
2
A total of 70 m of 16 mm
2
A total of 50 m of 2.5 mm
cable is included
cable is included
Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x