4.1.2

Recess Dimensions and Side Views for 12" Floor Boxes

Recess Positions and Dimensions
• *Cable conduit should ideally be a non-flexible PVC pipe with a minimum inner diameter of 100
mm and a maximum outer diameter of 130 mm.
• See page 39 for worst case positioning.
• After placing the box, pour concrete into the hole so that the aluminum cover rests on a
minimum of 10 mm stable concrete on all sides.
• The absolute minimum depth of the pit over the whole surface is 310 mm.
• The recess surface must be even for fine adjustment of the floor boxes.
Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x
Cable conduit*
min. 780
834
+/-7mm
Isocenter
Figure 11
Cable conduit*
1357
+/-7mm
37