CEILING AND WALL MOUNTS
5.5

Ceiling Mounted In-Room Distribution Box

General Information
The following tables provides information on the drilling pattern and required space for service for
the in-room distribution box:
Component
Dimensions
Weight
Safety Factor
Material Fire Resistance
Fire Protection Measures
Bolts
Location
Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x
572mm
Drilling pattern
In Room Distribution Box
Boreholes 6mm
Figure 32
Box
Top Cover
Service Area
Description
600 x 300 x 170 mm (length x width x depth)
8 kg (fully equipped)
Four (refer to local regulations)
UL94 V-2
According to local regulations
You must supply four bolts for ceiling fixation,
according to local regulations.
The in-room distribution box (1.3) should be in-
stalled so that the cable outlet (edge of the
box) is:
• Maximum 4 m away from the black interface
plate of the celling mounted monitor arm
(1.2) or
• Maximum 7 m away from the wall mount of
the wall mounted monitor arm (1.2).
• Maximum 4.5 m away from the external blue-
tooth module (3.2).
If cables cannot be routed directly, plan the dis-
tance between the two related devices respec-
tively shorter.