How to run the Self Test

Verification Manual
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Where dual wavelength optical transmitters fitted the self test should be
NO TE
re-cabled and re-run to verify the alternative transmitters wavelength
output. Select the cable configuration from the self test menu as
appropriate.
Safety precautions care and connection cleanliness must be observed
to ensure that the optical connections are not damaged or degraded.
Ensure the recommended optical attenuation is present in all optical
loopback connections as failure to do so could cause self test failure or
damage to the optical receivers.
1 Press <Menu>, choose System > Self Test then press
<Select>.
2 Ensure all loopbacks are in place and, from the self test page
menu, select the appropriate 1310/1550 nm 2.5 G - 52 M
optical loopback Cabling Configuration.
3 On the self test page choose Run Selected Test and choose
Confidence Test. Choose START to begin the test.
4 The self test will now run selecting each sub-test in sequence.
The remaining test time to complete the tests is displayed on
the right hand side of the self test page.
5 If any sub-test fails an error message and error code will be
returned. Up to five errors are recorded.
6 On the self test page choose Run Selected Test and choose
Frequency Measurement. Choose START to begin the test.
Further information on the self test can be obtained in the self
test overview section at the rear of this manual.