5.6.5 Installing the user interface certificate

The user interface of the product is protected by https encryption together with a password. A user interface certi­
ficate is installed for this purpose.
► NOTICE! Once a month, check the digital version of the instructions for use to see whether there are new
instructions in this "Installing the user interface certificate" section advising you to install a new user interfaces
certificate that maintains digital protection and access to the user interface.

5.7 Troubleshooting

The following subsections contain further information on eliminating errors:
"User interface error messages" (see Page 39)
"Fault table" (see Page 41)
"Measures" (see Page 43)
"Replacing faulty product components" (see Page 44)

5.7.1 User interface error messages

61
1234567.bionicpro.local
OK
User interface error messages are displayed in a warning window  .
The error message can be temporarily ignored by tapping the "Back"   browser button or selecting "Update
later"  . In this case, the red warning icon 
to them. The digit in the red circle indicates the number of ignored error messages. Tapping the red warning icon
displays the list 
of error messages which can in turn be accessed again and then processed by tapping them.

5.7.1.1 "No connection to sensor" error message

62
Take the following steps for the "No connection to sensor" error message:
► Check whether the sensor is switched on ("Charging switched off sensors in the base station and switching
them on prior to measuring" section – see Page 15).
► Check whether the battery level is too low ("System information" section – see Page 27).
► Once the sensor is charged and is detected by the base station, tap the "OK"   button.
743A250 Bionic Pro
100 %
Später aktualisieren
reminds you of the ignored error messages by calling your attention
100 %
1234567.bionicpro.local
Es konnte keine Verbindung zu Sensor X
hergestellt werden. Bitte prüfen Sie den
Ladestand der Batterie.
OK
100 %
Messung mit Orthese
3
3
Es konnte keine Verbindung zu Sensor X
hergestellt werden. Bitte prüfen Sie den
Ladestand der Batterie.
3
Akkustand niedrig
Firmware aktualisieren
Verbindungsfehler
OK
Use
39