17.2 Operating states/error signals

The prosthesis indicates operating states and error messages through beeps and vibration signals.

17.2.1 Signals for operating states

Battery charger connected/disconnected
Beep signal
1x short
Mode switching
INFORMATION
When mute mode (silent mode) is activated, no beep and vibration signals are generated.
INFORMATION
If the Volume parameter is set to '0' in the Cockpit app, there are no beep signals (see Page 93).
Beep signal
Vibration sig­
nal
1x short
1x short
1x long
1x long
1x short
1x short
2x short
2x short
3x short
3x short

17.2.2 Warnings/error signals

Error during use
Beep signal
5 x long
the
product
settings
Vibration signal
3x long
1x before beep signal
Additional action performed
Mode switching using the Cockpit
app
Bouncing on the forefoot followed
by unloading the prosthetic leg
Weight taken off prosthetic leg
and leg kept still for 1 second
Weight taken off prosthetic leg
and leg kept still for 1 second
Weight taken off prosthetic leg
and leg kept still for 1 second
Vibration signal
1 x long at intervals of
approx.
5 seconds
mute mode (silent mode)
is activated, this signal is
not generated)
3 x long
5 x long
5 x long repeated every
60 seconds
using
the
corresponding
Event
Charging mode started (3 sec. after connecting the bat­
tery charger)
Self-test completed successfully, product is operational
Event
Mode switching is performed using the
Cockpit app.
Bouncing pattern recognised.
Switching to basic mode (mode 1) car­
ried out.
Switching to MyMode 1 (mode 2) car­
ried out.
Switching to MyMode 2 (mode 3) car­
ried out.
Result
Overheated hydraulic unit Reduce activity.
(if
Charge level under 25%
Charge level under 10%
Medium
error
Page 100)
e.g. a sensor is not oper­
ational
adjustment
software
Required action
Charge battery soon.
Remaining
operating
approx. 24 hours
Charge battery soon
Remaining
operating
approx. 6 hours
(see
Walking possible with restric­
tions. Please note the change
in flexion resistance.
The product must be inspec­
ted immediately by an author­
ised Ottobock Service Center.