Information On The Use Of A Mobile Device With The Cockpit App; Scope Of Delivery And Accessories; Scope Of Delivery - Otto Bock Genium Series Instructions For Use Manual

Bionic prosthetic system
Table of Contents

Advertisement

Available languages

Available languages

Scope of Delivery and Accessories

4.11 Information on the use of a mobile device with the cockpit app

CAUTION
Improper use of the device
Falling due to altered damping behaviour as a result of unexpected switching into MyMode.
► Use the instructions for use (user) to instruct the patient on proper use of the device with the cockpit app.
CAUTION
Improper use of the setting parameters in the MyModes
Falling due to unexpected product behaviour because of changed damping behaviour.
► Instruct the patient regarding the functionality and adjustment options for all parameters of the MyModes.
CAUTION
Independently applied changes or modifications made to the device
Falling due to altered damping behaviour as a result of unexpected switching into MyMode.
► Do not make any independent changes to the hardware of the device.
► Do not make any independent changes to the software/firmware of the device which are not included in the
update function of the software/firmware.
CAUTION
Improper mode switching with the device
Falling due to unexpected product behaviour because of changed damping behaviour.
► Ensure that the patient stands securely during all switching processes.
► Inform the patient that the changed damping characteristics have to be verified after switching, and feedback
from the acoustic signal emitter and the device display must be observed.
► Switching back to basic mode is mandatory once the activities in MyMode have been completed.
NOTICE
Destruction of the device due to falling or penetration of water
Malfunction of the device.
► Note the instructions for use for the device.
► If it should no longer be possible to switch back from a MyMode to basic mode, the component can only be
switched back to basic mode by using a movement pattern (see Page 99) or by connecting/disconnecting the
inductive battery charger.
NOTICE
Failure to observe the system requirements for the installation of the cockpit app
Malfunction of the device.
► Install the cockpit app only on the operating systems listed in the section "System requirements" (see
Page 86).
INFORMATION
The illustrations in these instructions for use are only examples and may deviate from the respective mobile device
being used and the version.
5 Scope of Delivery and Accessories

5.1 Scope of delivery

1 pc. 3B1=ST Genium - Bionic Prosthetic System
(with threaded connector) or
1 pc. 3B1 Genium - Bionic Prosthetic System (with
pyramid connector) or
78
1 pc.
3B1-2=ST/3B1-2=ST-9.2
threaded connector) or
1 pc. 3B1-2/3B1-2=9.2 Genium (with pyramid con­
nector)
1 pc. 2R20 AXON tube adapter or
1 pc. 2R21 AXON tube adapter with torsion
Genium
(with

Advertisement

Chapters

Table of Contents
loading

This manual is also suitable for:

Genium 3b1/3b1-st3b1-2/3b1-2-st

Table of Contents