Notes On The Safety Modes; Instructions For Use With An Osseointegrated Implant System - Otto Bock Genium Series Instructions For Use Manual

Bionic prosthetic system
Table of Contents

Advertisement

Available languages

Available languages

CAUTION
Quickly pushing the hip forward with the prosthesis extended (e.g. serve while playing tennis)
>
Falling due to unexpected activation of the swing phase.
► Note that the knee joint may flex unexpectedly when the hip is pushed forward quickly while the prosthesis is
extended.
► If the patient participates in sports where this movement pattern can occur, configure corresponding
MyModes using the adjustment software. For further information about the MyModes, see the section
'MyModes' (see Page 97).

4.9 Notes on the safety modes

CAUTION
Using the product in safety mode
Falling due to unexpected product behaviour because of changed damping behaviour.
► The warnings/error signals (see Page 107) have to be observed.
► Particular caution is necessary when using a bicycle without a freewheel (with a fixed gear).
CAUTION
Safety mode cannot be activated due to malfunction caused by water penetration or mechanical damage
Falling due to unexpected product behaviour because of changed damping behaviour.
► Using the product when it is defective is prohibited.
► The product must be inspected by an authorised Ottobock Service Centre.
CAUTION
Safety mode cannot be deactivated
Falling due to unexpected product behaviour because of changed damping behaviour.
► If safety mode cannot be deactivated by recharging the battery, a permanent error has occurred.
► Using the product when it is defective is prohibited.
► The product must be inspected by an authorised Ottobock Service Centre.
CAUTION
Safety signal occurs (ongoing vibration)
Falling due to unexpected product behaviour because of changed damping behaviour.
► The warnings/error signals (see Page 107) have to be observed.
► After the safety signal has been emitted, further use of the product is prohibited.
► The product must be inspected by an authorised Ottobock Service Centre.

4.10 Instructions for use with an osseointegrated implant system

Only valid for the products 3B1-2/3B1-2=ST
WARNING
High mechanical loads due to normal or unusual situations, such as falling
>
Overloading of the bone, which can lead to pain, loosening of the implant, necrosis or fracture among other
things.
>
Damage or breakage of the implant system or its components (safety components...).
► Verify compliance with the fields of application, conditions of use and indications according to the information
of the manufacturers, both for the knee joint and for the implant system.
► Note the instructions of the clinical personnel that indicated the use of the osseointegrated implant system.
Safety
77

Advertisement

Chapters

Table of Contents
loading

This manual is also suitable for:

Genium 3b1/3b1-st3b1-2/3b1-2-st

Table of Contents