9.

1 Messages

1.1 Pending Errors
1.1.5
1.2 Maintenance List
1.2.5
1.3 Message List
1.3.1

2 Diagnostics

2.1 Identification
2.1.3
2.1.4
2.2 Sensors
2.2.1
A-96.150.541 / 111223
Program List and Explanations
Provides the list of active errors with their status (active, acknowl-
edged). If an active error is acknowledged, the alarm relay is active
again. Cleared errors are moved to the message list.
Provides the list of necessary maintenance. Cleared maintenance
messages are moved to the Message list.
Shows the error history: Error code, date / time of issue and status
(active, acknowledged, cleared). 65 errors are memorized. Then the
oldest error is cleared to save the newest error (circular buffer).
In diagnostics mode, the values can only be viewed, not modified.
Desig.: Designation of the instrument.
Version: Firmware of instrument (e.g. 1.00-11/23)
Factory Test: Test date of the Instrument and Motherboard.
Operating Time: Shows the operating time in Years, Days, Hours,
Minutes and Seconds.
Oxytrace G
Current value: Shows the measured value in ppb.
Shows the temperature-compensated measured
Raw value tc:
value in mA.
Shows the saturation in %
Saturation