Maintenance checklist

ü Inspect the H5i water tub and flip lid seal for wear and deterioration.
ü Replace the water tub if any component is leaking or has become cracked, cloudy or pitted.
ü Replace the flip lid seal if cracked or torn.
ü Clean white powder deposits in the water tub by using a solution of one part household vinegar to 10
parts water.

Replacing the air filter

Replace the air filter every six months (or more often if necessary).
1. Remove the air filter cover from the back of the VPAP device.
2. Remove and discard the old filter.
3. Insert a new ResMed air filter ensuring that it is sitting flat in the air filter cover.
4. Replace the air filter cover.
Air filter cover
Air filter
The following filters are available for use with VPAP devices:
Filter
Standard (ASMB 160)
Hypo-allergenic (Air Safety Electret100 – electrostatic filter)

Antibacterial filters

Antibacterial filters are not compatible with the ClimateLine or SlimLine tubing but can be used with the
Standard or 3m air tubing.
Antibacterial filters increase resistance in the air circuit and may affect accuracy of displayed and
delivered pressure, particularly at high flows. ResMed recommends using a filter with a low impedance
(eg, less than 2 cm H
30
O at 60 L/min).
2
Efficiency
88% at 7 micron
89.8% at 0.5 micron, bacterial
efficiency of 99.54%.