Maximum Pressure Test; Flow Calibration Test; Blower Test; Keypad Test - ResMed VPAP Auto 25 Service Manual

Hide thumbs Also See for VPAP Auto 25:
Table of Contents

Advertisement

10 Maximum pressure test

1 Select Maximum pressure test in the Xplorer bar, or click on the Perform the Maximum
pressure test icon
2 The test set-up remains the same as the pressure calibration test.
3 Select Next and follow the steps as instructed.
If the flow generator fails this test, follow the procedures outlined in Section 7 Calibration
procedures to do a full recalibration of the flow generator, ensuring the equipment set-up is
correct.

11 Flow calibration test

1 Select Flow calibration test in the Xplorer bar, or click on te Perform the flow calibration test icon
.
2 Set up the test equipment as instructed by the test software.
3 Select Next and follow the steps as instructed.
If the flow generator fails this test, follow the procedures outlined in Section 7 Calibration
procedures to do a full recalibration of the flow generator, ensuring the equipment set-up is
correct.

12 Blower test

1 Select Blower test in the XPlorer bar, or click on the Perform the blower test icon
2 Set up the test equipment as instructed by the test software.
3 Select Next and follow the steps as instructed.
If the flow generator fails this test, repeat the pressure calibration, ensuring the test set-up is
correct. If the unit continues to fail, refer to Section 7 Troubleshooting guide.

13 Keypad test

1 Select Keypad test in the Xplorer bar, or click on te Perform the keypad test icon
2 The test set-up remains the same as the blower test.
3 Select Next and follow the steps as instructed.
If the flow generator fails this test, refer to Section 8 Troubleshooting guide.

14 LCD test

1 Select LCD test in the XPlorer bar, or click on the Perform the LCD test icon
2 Follow the steps as instructed.
If the flow generator fails this test, refer to Section 7 Troubleshooting guide.

15 Audible alarm test

1 Select Audible alarm test in test in the Xplorer bar, or click on te Perform the audible alarm test
icon
2 Set up the tset equipment as instructed by the test software.
3 Select Next and follow the steps as instructed.
If the flow generator fails this test, refer to Section 8 Troubleshooting guide.
VPAP™
14
Service Manual
.
.
.
.
.