Overview

Interfaces

A There are 2 ports on the rear of Fabius for
communication with external devices. The
ports are marked with COM 1 and COM 2
These ports are used for data communication
using the Dräger MEDIBUS or Vitalink data
protocols.
WARNING
Risk of electric shock
Connecting devices to the MEDIBUS
interfaces can lead to an increased leakage
current. If the protective ground of one of
these devices fails, the leakage current may
rise above the permissible values.
– Only connect with the approval of the
respective device manufacturer.
– Have the leakage current checked by
service personnel.
– If the permissible value is exceeded,
disconnect the devices from the MEDIBUS
interface.
*
optional
32
CAUTION
Risk of electrical safety
To ensure electrical safety, only connect devices
to the serial ports (COM 1 and COM 2) with a
A
maximum nominal voltage of 24 Vdc that meet
one of the following standards:
– IEC 60950-1: ungrounded SELV circuits
– IEC 60601-1 (as of 2nd edition): exposed
secondary circuits
CAUTION
Risk of device failure
Only use monitors, mounting parts, and
connection cables that have been approved by
Dräger.
NOTE
The data transmission of the gas analysis data
must be activated by DrägerService or the
*
authorized local service partner.
.
Instructions for use Fabius Tiro SW 3.n