Interfaces And Ports - Dräger Atlan A300 Instructions For Use Manual

Hide thumbs Also See for Atlan A300:
Table of Contents

Advertisement

16.18
Operating characteristics (continued)
Weight of the large version
Nominal configuration consisting of mechani-
cally controlled gas mixer, plug-in connector for
3 vaporizers, breathing system, CLIC adapter
and CLIC absorber, breathing hoses, central
supply hoses (5 m (16.4 ft)), scavenging hose
(5 m (16.4 ft))
Permissible total weight
Touchscreen color screen
Screen diagonal
Background illumination
Resolution
RFID system
Operating frequency
Transmitter power
Modulation
Electromagnetic compatibility
Protection classes
Device
Applied parts (connections for breathing hoses)
Degree of protection
Classification in compliance with Directive
93/42/EEC, Annex IX
UMDNS code
Use of latex
16.19

Interfaces and ports

Serial ports
Information
|
Instructions for use
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Approx. 160 kg (353 lbs)
330 kg (727 lbs)
Approx. 39 cm (15.3 in)
LED
1280 x 768 pixels
13.56 MHz ± 50 ppm (wideband)
≤42 dBμA/m (200 mW ± 1 dB)
ASK (amplitude shift keying)
Tested in compliance with IEC 60601-1-2
I, in compliance with IEC 60601-1
TYPE BF
IP20 according to IEC 60529,
meets ISO 80601-2-13
II b
10-134
The device is made without natural rubber
latex.
COM 1 and COM 2
Only connect devices that meet the require-
ments of IEC 60950-1 for ungrounded SELV
circuits and the requirements of IEC 60601-1
(as of the 2nd edition) for touchable secondary
circuits with a maximum nominal voltage of
24 V DC.
Use COM2 to connect devices for external
patient-gas measurement with the MEDIBUS
communication protocol.
Technical data
287