Performing User Calibration - Agilent Technologies SPS 4 User Manual

Autosampler
Table of Contents

Advertisement

7
Running the Calibration Wizard

Performing User Calibration

To perform user calibration:
1 Ensure the autosampler power is on.
Close the instrument software before launching the Calibration Wizard.
N O T E
2 Start the Calibration Wizard software. The Calibration Wizard window
3 Select the appropriate COM port from the drop- down list, and then
4 If the calibration wizard can connect to your autosampler successfully,
74
The sample racks fitted should be empty and not contain any sample
vials. If a microtiter plate Adapter is fitted, it should be loaded with
four 96 well microtiter plates.
opens as shown below.
click Connect in the Com Port field.
If you chose the wrong one, an error message appears. If the
appropriate COM port number is not listed, check the autosampler
power, the USB connection, and the COM port setting in the instrument
control software.
the field under the Com Port field shows the information on the rack
layout specific to your autosampler. Ensure that the sample rack types
displayed on the screen correspond to those loaded into their respective
positions on the rack location mat.
Agilent SPS 4 Autosampler User's Guide