the peristaltic pump, as shown in
Figure 17

Loading Wash Pump Tubing

1 Release the pressure plate by disengaging the occlusion arm.
2 Fit one of the outside bridges on the pump tubing into the appropriate
3 Feed the tubing around the rotor and fit the center tube bridge into the
4 Reposition the pressure plate over the rotor and fix by slotting the
5 Route the outlet of the wash pump tubing through the tube anchor clips
6 Route the inlet line of the wash pump tubing through the tube anchor
Agilent SPS 4 Autosampler User's Guide
Wash Reservoir
Flow
Pumped Drain Arrangement
left tube clamping block.
right tube clamping block. Ensure there are no twists or kinks in the
tubing after loading as this will adversely affect tube life.
occlusion arm back into the track slot.
to the delivery line or input of the fixed wash reservoir.
clip(s) and connect this to the suction line from the rinse solution
reservoir.
Figure
17.
Wash Pump
FWD
ROTATION
Waste Out
Installation