Fitting The Sample Probe - Agilent Technologies SPS 4 User Manual

Autosampler
Table of Contents

Advertisement

3
Installation

Fitting the Sample Probe

Always switch off the power to the autosampler before installing or removing the
WA R N I N G
sample probe.
Z-Axis Slide
Eyelet for tube routing
Tube routing this side
Figure 13
To fit the sample probe:
1 Turn the autosampler off.
2 Manually rotate the probe arm so it faces the front of the autosampler
3 Undo the mount nut from the probe mount block on the sample probe
4 Place the mount nut from the sample probe on top of the accessories
5 Move the Z- axis slide down stopping just above mount nut sitting on
42
Installing the sample probe
for easy access.
and slide the nut off the sample probe.
plate with the recessed edge facing up and the holes approximately
aligned.
the accessories plate.
Tube restraint
Probe mount block
Mount nut
Sample probe
Accessories plate
Agilent SPS 4 Autosampler User's Guide