Installation
Bench Requirements
The autosampler should be placed on a bench with a height between
790 mm (31.1") and 910 mm (35.8") for optimal user ergonomics. A
footprint of 600 mm
should be flat with no more than 3 mm (7/64") variation. If a sampler
cover kit is fitted, additional space should be allowed on the side where
the exhaust hose ducting is located.
The autosampler should be located in close proximity to your instrument
so as to reduce the delay caused by the length of capillary tubing
connecting to the autosampler.
Inspection and Unpacking
Unpacking and Lifting
Keep the shipping box and packaging. If you need to return the autosampler during the
N O T E
warranty period, it is a requirement of the warranty that the equipment is repacked in the
original shipping box. For instructions on packing the equipment, refer to the
Packing/Unpacking Guide.
Inspect all items for damage. Any damage or missing items should be reported to
C A U T I O N
Agilent immediately. Cite the part number and serial number of your autosampler.
Danger to hands and feet. When unpacked the autosampler is heavy. To avoid injury
WA R N I N G
to personnel or damage to equipment, always use at least two people to lift the
autosampler into position.
32
×
×
363 mm (23.6"
14.3") is required. The bench
Agilent SPS 4 Autosampler User's Guide