Initialization Problems

The initialization sequence is a powerful self- diagnostic tool. It runs
immediately after the autosampler is powered on and tests the
functionality of the autosampler. The probe arm goes through a sequence
of movements and stops above the fixed wash reservoir position on the
rack location mat. If there is a problem with the autosampler, then it will
likely manifest itself at some stage during the initialization sequence.
The correct initialization sequence can be seen on
If the autosampler fails to complete initialization, the power/error
indicator illuminates to display the error condition. See
error indicator codes.
Check the Agilent SPS 4 by doing the following:
1 Turn the power off and move each axis manually until they stop and
2 If no mechanical obstruction can be identified turn the unit back on
3 If unsuccessful, contact your Agilent customer service representative.
68
Error Indicator Codes
E2
E3
E4
Error
Code
R
0
R
13
R
R
0
14
R
R
R
15
determine if there is any mechanical obstruction preventing full travel.
and allow it to attempt the initialization process again.
Error Description
Z-axis movement detected without the
motor being commanded to move.
Reserved.
CPU exception error
Agilent SPS 4 Autosampler User's Guide
Action
Refer to
"Probing
Problems"
on Page 69
for details.
Reinitialize the SPS 4. If
error persists, contact
your authorized Agilent
representative.
Page
50.
Table 7