Installation
Connecting ICP Expert software to the ICP-OES
To connect the software and instrument:
1
2
3
4
5
6
Performing a Detector and Wavelength Calibration
To perform a detector and wavelength calibration:
1
2
3
4
Once the detector calibration is completed, the date and time of the last successful
NOTE
calibration will be displayed, confirming the detector measurement has been
completed.
5
6
24
Start the ICP Expert software by double-clicking on the desktop icon.
Click Instrument.
Click Connect.
Enter the IP address of the instrument or select an existing instrument.
Click Connect.
Click Close.
Turn on the external exhaust.
Start the ICP Expert software and open the Instrument Window.
Connect the software to the instrument, if needed.
a
Click Instrument in the ICP Expert toolbar.
b
Click Connect.
c
Select the instrument from the list and the click Connect.
Click Calibrate in the 'Detector' section.
Ensure a standard glass concentric nebulizer, a double pass spray chamber
and the appropriate plasma torch (for radial or dual view) are installed. Use
manual sampling. The recommended peristaltic pump tubing is White/White
for the pump and Blue/Blue for the drain.
Ignite the plasma.
Agilent 5800 and 5900 ICP-OES User's Guide