Error List - Agilent Technologies SPS 4 User Manual

Autosampler
Table of Contents

Advertisement

10 When all four points have been captured, a confirmation message is
11 After calibration has been successfully completed, check the positioning

Error List

On running the calibration wizard software, you may encounter any error
listed below:
• Serial Port Error
Agilent SPS 4 Autosampler User's Guide
displayed asking if you want to write the new calibration values.
• Click Yes to write the calibration values.
The autosampler is adjusted using the new calibration value.
• Click No if you want to retry.
The calibration process is restarted.
of the sample probe by clicking on each corner position button, and
driving the Z- Axis position down. Confirm that the sample probe is
positioned in the center of each well.
Example: Autosampler did not respond.
A serial port error occurs if the software cannot communicate with the
autosampler. Check that the power is on, and no error lights are
illuminated. Also confirm the USB connection and the COM port setting
in the instrument control software. If the problem still remains
unresolved, contact your Agilent customer support representative.
Running the Calibration Wizard
7
79

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents