Table 7
E1
0
0
0
0
R
R
R
R
R
Agilent SPS 4 Autosampler User's Guide
Error Indicator Codes
E2
E3
E4
Error
Code
R
0
0
4
R
0
R
5
R
R
0
6
R
R
R
7
0
0
0
8
0
0
R
9
0
R
0
10
0
R
R
11
R
0
0
12
Error Description
RAM test error
Axis initialization error
X-axis position error. The X-axis motor
was commanded to move, but did not
reach the desired position.
Theta-axis position error. The
Theta-axis motor was commanded to
move, but did not reach the desired
position.
Z-axis position error. The Z-axis motor
was commanded to move, but did not
reach the desired position.
EEPROM error
Mat checksum error
X-axis movement detected without the
motor being commanded to move.
Theta-axis movement detected without
the motor being commanded to move.
Troubleshooting
6
Action
Reinitialize the SPS 4. If
error persists, contact
your authorized Agilent
representative.
Reinitialize the SPS 4. If
error persists, contact
your authorized Agilent
representative.
Refer to
"Probing
Problems"
on Page 69
for details.
Refer to
"Probing
Problems"
on Page 69
for details.
Refer to
"Probing
Problems"
on Page 69
for details.
Reinitialize the SPS 4. If
error persists, contact
your authorized Agilent
representative.
Reinitialize the SPS 4. If
error persists, contact
your authorized Agilent
representative.
Refer to
"Probing
Problems"
on Page 69
for details.
Refer to
"Probing
Problems"
on Page 69
for details.