Agilent Technologies SPS 4 User Manual page 45

Autosampler
Table of Contents

Advertisement

The wash pump accommodates standard 3 bridged pump tubing which is
available in a range of internal diameters for delivering different flow
rates. The optimal distance between the bridges on the pump tubing is
82
Wash pump tubing is available in different tube materials so ensure that you select a tube
N O T E
material that is compatible with the rinse solution. If you are not sure which material to use,
contact your Agilent representative for a recommendation.
When a pump channel is not in use, release the occlusion arm on the channel and remove
the tubing from the guides to extend the life of the tubing.
Figure 15
To get access to the wash pump to fit or change the pump tubing, the
autosampler may need to be moved.
Before moving the autosampler, ensure all sample racks and the standards rack have been
N O T E
removed. In addition it may be necessary to disconnect the power cable, interface cable,
sample line, inlet line to the wash pump or the drain line from the fixed wash reservoir
depending on the distance moved.
Agilent SPS 4 Autosampler User's Guide
85 mm.
Pressure plate
Rotor
Tube anchor clip
Wash pump tube routing and connections
Installation
Occlusion arm and spring
Tube clamping block
Pump tubing
45