To change the sample probe or access the fixed wash reservoir, it is
recommended to first remove the front cover. Slide the front cover up
until the handle is almost touching the chassis and then flex it slightly
forward to remove.

Connections

General
There is an entry/exit point to the fixed wash reservoir through a rubber
grommet on the rear cover.
Exhaust Ducting
On the side cover there is a ducting connection for connecting exhaust
ducting. The hole will suit ducting with an internal diameter of 50 mm
(1.97") used in conjunction with a cable tie or hose clamp.
Care and Maintenance
• Do not use abrasive cleaners or strong solvents to clean the cover as
• Use a soft cloth and household window cleaner to clean off fingerprints
• If using volatile chemicals ensure the cover is adequately evacuated at
It is strongly recommended that an autosampler cover is fitted to the autosampler
WA R N I N G
when it is used with samples, standards or rinse solution(s) that release corrosive
fumes in any concentration. Whenever such liquids are present, the atmospheric
contents of the sample cover must be extracted continuously at a minimum rate of
350 liters per minute. Corrosive samples and standards should be removed from the
autosampler once analysis is completed.
92
this will reduce the clarity of the cover.
on the cover.
all times.
Agilent SPS 4 Autosampler User's Guide