:POWer:PSRR:WGEN:LOAD

Command Syntax
:POWer:PSRR:WGEN:LOAD <impedance>
<impedance> ::= {ONEMeg | FIFTy}
The :POWer:PSRR:WGEN:LOAD command sets the waveform generator expected
output load impedance.
The output impedance of the Gen Out signal is fixed at 50 ohms. However, the
output load selection lets the waveform generator display the correct amplitude
and offset levels for the expected output load. If the actual load impedance is
different than the selected value, the displayed amplitude and offset levels will be
incorrect.
Query Syntax
:POWer:PSRR:WGEN:LOAD?
The :POWer:PSRR:WGEN:LOAD? query returns the waveform generator expected
output load impedance setting.
Return Format
<impedance><NL>
<impedance> ::= {ONEM | FIFT}
See Also
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
(see
page
1354)
":POWer:PSRR"
on page 632
":POWer:PSRR:APPLy"
":POWer:PSRR:DATA"
":POWer:PSRR:FREQuency:MAXimum"
":POWer:PSRR:FREQuency:MINimum"
":POWer:PSRR:FREQuency:MODE"
":POWer:PSRR:PPDecade"
":POWer:PSRR:SOURce:INPut"
":POWer:PSRR:SOURce:OUTPut"
":POWer:PSRR:WGEN:VOLTage"
":POWer:PSRR:WGEN:VOLTage:PROFile"
on page 633
on page 634
on page 635
on page 636
on page 637
on page 639
on page 640
on page 641
on page 644
on page 645
24
:POWer Commands
643

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the InfiniiVision M9241A and is the answer not in the manual?

Subscribe to Our Youtube Channel

This manual is also suitable for:

Infiniivision m9242aInfiniivision m9243a

Table of Contents