Trigger:pxi:mslot - Keysight InfiniiVision M9241A Scpi Programmer's Manual

Pxie oscilloscopes
Hide thumbs Also See for InfiniiVision M9241A:
Table of Contents

Advertisement

31
:TRIGger Commands

:TRIGger:PXI:MSLot

Command Syntax
:TRIGger:PXI:MSLot <slot_number>
<slot_number> ::= 2 to (# chassis slots) in NR1 format
When this PXIe oscilloscope is a slave, the :TRIGger:PXI:MSLot command specifies
the chassis slot number of the master module.
It is important to identify the master slot number on each slave so that the proper
set of calibration factors can be used when aligning waveforms.
Calibration factors are created and saved using the KtScopeCal.exe application on
the chassis controller.
Query Syntax
:TRIGger:PXI:MSLot?
The :TRIGger:PXI:MSLot? query returns the master slot number setting.
Return Format
<slot_number><NL>
<slot_number> ::= 2 to (# chassis slots) in NR1 format
See Also
1120
(see
page
1354)
":TRIGger:PXI:MALine<n>:ENABle"
":TRIGger:PXI:MODE"
":TRIGger:PXI:SALine"
":TRIGger:PXI:TLINe"
on page 1123
":ACQuire:RSIGnal"
on page 250
":TRIGger:PXI:SYNC"
on page 1122
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
on page 1118
on page 1119
on page 1121