:TRIGger:PXI:TLINe

Command Syntax
:TRIGger:PXI:TLINe <trigger_line>
<trigger_line> ::= {PTRig<n>}
<n> ::= 0 to (# chassis lines - 1) in NR1 format
The :TRIGger:PXI:TLINe command specifies the chassis PXI_TRIG line (0-7) to be
used for the multi-module trigger signal.
The master module and all slave modules must use the same trigger line.
If the modules are in different chassis trigger bus segments (which are isolated by
default), you must use the chassis Soft Front Panel (SFP) application to connect
the selected PXI_TRIG line between trigger bus segments.
Query Syntax
:TRIGger:PXI:TLINe?
The :TRIGger:PXI:TLINe? query returns the specified trigger line.
Return Format
<trigger_line><NL>
<trigger_line> ::= {PTR<n>}
See Also
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
(see
page
1354)