:SBUS<n>:SENT:TRIGger:TOLerance
Command Syntax
:SBUS<n>:SENT:TRIGger:TOLerance <percent>
<percent> ::= from 1-28, in NR1 format.
The :SBUS<n>:SENT:TRIGger:TOLerance command specifies the tolerance
variation that is considered a violation.
The trigger tolerance can be up to the :SBUS<n>:SENT:TOLerance setting minus
two percent. For example, after sending ":SBUS1:SENT:TOLerance 20", the
:SBUS1:SENT:TRIGger:TOLerance setting can be from 1 to 18.
Query Syntax
:SBUS<n>:SENT:TRIGger:TOLerance?
The :SBUS<n>:SENT:TRIGger:TOLerance? query returns tolerance variation
percent setting.
Return Format
<percent><NL>
<percent> ::= from 1-28, in NR1 format.
See Also
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
(see
page
1354)
":SBUS<n>:SENT:CLOCk"
":SBUS<n>:SENT:CRC"
":SBUS<n>:SENT:DISPlay"
":SBUS<n>:SENT:FORMat"
":SBUS<n>:SENT:IDLE"
":SBUS<n>:SENT:LENGth"
":SBUS<n>:SENT:PPULse"
":SBUS<n>:SENT:SIGNal<s>:DISPlay"
":SBUS<n>:SENT:SIGNal<s>:LENGth"
":SBUS<n>:SENT:SIGNal<s>:MULTiplier"
":SBUS<n>:SENT:SIGNal<s>:OFFSet"
":SBUS<n>:SENT:SIGNal<s>:ORDer"
":SBUS<n>:SENT:SIGNal<s>:STARt"
":SBUS<n>:SENT:SOURce"
":SBUS<n>:SENT:TOLerance"
":SBUS<n>:SENT:TRIGger"
":SBUS<n>:SENT:TRIGger:FAST:DATA"
":SBUS<n>:SENT:TRIGger:SLOW:DATA"
":SBUS<n>:SENT:TRIGger:SLOW:ID"
":SBUS<n>:SENT:TRIGger:SLOW:ILENgth"
on page 868
on page 869
on page 870
on page 872
on page 874
on page 875
on page 876
on page 878
on page 879
on page 881
on page 883
on page 885
on page 887
on page 889
on page 890
on page 891
on page 893
on page 894
on page 896
on page 898
27
:SBUS<n> Commands
899
Need help?
Do you have a question about the InfiniiVision M9241A and is the answer not in the manual?