Press the <Right Arrow> to move into the [Channel Setup] menu as shown below. Note that the
highlight is on the first menu cell: [Recall Site].
Use the <Down Arrow> to move the highlight to [Create/Name Site]. Create a site named: SITE1
(as detailed in paragraph 2.8.1).

2.9.2 HOW TO ENTER PIPE DATA

After creating the site, the next step is to enter the Pipe Data. The Pipe Data menu allows you to either
enter the pipe data manually or select a preset pipe configuration from an option list. This section
explains how to enter the data manually. The two mandatory entries to be entered in this menu are the
pipe's Outer Diameter (OD) and its Wall Thickness. Here we will insert a pipe outer diameter of 9.00"
and a pipe wall thickness of 0.29" as an example.
Press the <Left Arrow> to leave [Channel Setup]. Press the <Down Arrow> and then the <Right
Arrow> to enter the [Pipe Data] menu.
Move the highlight to [Pipe OD <in>] and then press the <Right Arrow> and then the <Right Arrow>.
Note that an equal sign appears before the current parameter.
Siemens
Site
Create-Name-Recall-Enable & Delete Site
Channel Setup
Pipe Data
Application Data
Pick/Install Xdcr
Operation Adjust
Flow/Total Units
DataSpan/Set/Cal
Datalogger Setup
I/O Data Control
Diagnostic Data